Metal wear particle characterization from metal on metal total hip replacements: transmission electron microscopy study of periprosthetic tissues and isolated particles

J Biomed Mater Res. 1998 Oct;42(1):103-11. doi: 10.1002/(sici)1097-4636(199810)42:1<103::aid-jbm13>3.0.co;2-m.

Abstract

The less intense tissue reaction around metal on metal total hip replacements (THRs) compared to metal on polyethylene (PE) THRs may be explained by the differences in the characteristics of metal wear particles. In this study, transmission electron microscopy was used to study metal wear particles that were either in situ in cells or had been extracted from the cells by a new technique based on enzymatic tissue digestion. The tissues were obtained from 13 patients undergoing revision of metal on metal THRs with cobalt-chromium-molybdenum (CoCrMo) bearing couples. Most of the CoCrMo wear particles were smaller than 50 nm (range 6-834 nm) and round to oval in shape with irregular boundaries. This size range is considerably smaller than that reported for PE particles. While even a small volume of metal wear will produce high numbers of particles, the apparently less severe local tissue reaction to metal particles may be due to the possibility that corrosion, dissolution, and dissemination of metal particles may result in fewer local biological effects than the long-term retention of PE particles in the periprosthetic tissues.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Biocompatible Materials
  • Electron Probe Microanalysis
  • Hip Prosthesis*
  • Humans
  • Macrophages / immunology
  • Metals / chemistry*
  • Microscopy, Electron
  • Middle Aged
  • Particle Size

Substances

  • Biocompatible Materials
  • Metals