Determination of dimethindene in human tears by high-performance liquid chromatography

J Chromatogr B Biomed Sci Appl. 1998 Apr 10;707(1-2):334-7. doi: 10.1016/s0378-4347(97)00589-6.

Abstract

A high-performance liquid chromatographic method is described for the determination of dimethindene in human tears. The tear samples were diluted in a 0.01 M hydrochloric acid-n-propanol mixture to prevent the irreversible adsorption of dimethindene. The diluted samples were directly injected into the chromatographic system to avoid sample pretreatment. The validation data demonstrate that the method is specific, precise and accurate within the calibration range of 12 to 1000 ng/ml dimethindene free base.