AutoCyte SCREEN interactive automated primary cytology screening system. A preliminary evaluation

Acta Cytol. 1998 Jan-Feb;42(1):185-8. doi: 10.1159/000331544.

Abstract

Objective: To perform a preliminary evaluation of the relative sensitivity of a new interactive automated cytology screening system (AutoCyte SCREEN, AutoCyte, Inc., Elon College, North Carolina, U.S.A.) designed as a primary screening system for AutoCyte PREP monolayer slides (AutoCyte).

Study design: Monolayer slides were prepared and evaluated in a masked review both manually and using the automated system. Human and machine performance was monitored and measured.

Results: Five hundred eighty-three monolayer preparations, including 56 clearly abnormal cases, were evaluated. The dual and combined computer and cytologist image evaluation resulted in a relatively low false positive rate, 20.8%, and 1.8% false negative rate. Use of the system was efficient, significantly reducing cytotechnologist screening time.

Conclusion: The integrated AutoCyte cell preparation and automated screening approach seems to hold significant promise for primary cytology screening applications.

Publication types

  • Comparative Study

MeSH terms

  • Adenocarcinoma / diagnosis
  • Adenocarcinoma / pathology
  • Artifacts
  • Automation
  • Carcinoma, Squamous Cell / pathology
  • Cell Biology
  • Cervix Uteri / cytology*
  • Clinical Laboratory Techniques
  • Diagnosis, Computer-Assisted / instrumentation*
  • Epithelial Cells / pathology
  • Equipment Design
  • Evaluation Studies as Topic
  • False Negative Reactions
  • False Positive Reactions
  • Female
  • Humans
  • Man-Machine Systems*
  • Mass Screening / instrumentation*
  • Microscopy
  • Observer Variation
  • Pilot Projects
  • Sensitivity and Specificity
  • Uterine Cervical Dysplasia / pathology
  • Uterine Cervical Neoplasms / pathology
  • Uterine Cervical Neoplasms / prevention & control*
  • Vaginal Smears / instrumentation*