Quantitative coronary angiographic and intravascular ultrasound assessment of a new nonarticulated stent: report from the Advanced Cardiovascular Systems MultiLink stent pilot study

J Am Coll Cardiol. 1998 Jan;31(1):50-6. doi: 10.1016/s0735-1097(97)00426-9.

Abstract

Objectives: The purpose of this study was to evaluate the safety, feasibility, optimal deployment technique and 1-year clinical outcome for the Advanced Cardiovascular Systems (ACS) MultiLink stent.

Background: Optimal stent deployment assessed by quantitative coronary angiography and intravascular ultrasound (IVUS) is associated with improved clinical outcome.

Methods: Forty-nine consecutive patients with a discrete stenosis in a native coronary artery 3 to 4 mm in diameter were treated with the new, balloon-expandable ACS MultiLink stent. Stent expansion was assessed in all patients using quantitative coronary angiography and serial IVUS imaging after 8-, 12- and 16-atm inflations. Clinical follow-up was obtained at 30 days and 1 year.

Results: All 49 patients had successful placement of a MultiLink stent without death, emergency coronary artery bypass graft surgery or Q wave myocardial infarction. After placement of the MultiLink stent, the minimal lumen diameter increased from 1.24 to 2.98 mm (p < 0.001), and diameter stenosis decreased from 61% to 7% (p = 0.001). Minimal lumen cross-sectional area by IVUS increased progressively after 8, 12 and 16 atm (5.6 to 6.8 to 7.4 mm2, respectively, p < 0.001). However, only 64% of stents achieved a lumen/reference area ratio > or = 70%. No adverse clinical events occurred by 30 days, and by 1 year only one patient (2.0%) required revascularization of the stented artery.

Conclusions: Treatment of stenoses in native coronary arteries with the MultiLink stent is associated with a high success rate and a low incidence of adverse events by 1 year, despite the fact that the majority of stents did not meet IVUS-defined criteria for "optimal stenting" derived from first-generation devices.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Coronary Angiography*
  • Coronary Disease / therapy*
  • Equipment Design
  • Feasibility Studies
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pilot Projects
  • Stents*
  • Treatment Outcome
  • Ultrasonography, Interventional*