Construction and performance of a photobleaching recovery apparatus with microsecond time resolution

Biophys Chem. 1996 Mar 7;59(1-2):203-10. doi: 10.1016/0301-4622(95)00139-5.

Abstract

A fluorescence recovery after photobleaching (FRAP) apparatus was constructed to measure sub-millisecond fluorescence recovery processes in living cells. The photobleaching pulse and probe beams were generated by modulating the intensity of a continuous wave Argon laser (4 W) by two acousto-optic modulators in series. The maximum intensity modulation was > 10(6):1 with a rise time of < 1 microsecond and a minimum pulse width of 6 microseconds. Fluorescence was detected by a photomultiplier, amplified by a transimpedance amplifier, and digitized at 1 MHz. During the photobleaching pulse, the photomultiplier gain was reduced by ca. 1500-fold by switching the second dynode voltage ca. 100 V negative with respect to the cathode voltage by computer control of two bidirectional Mosfet optoisolators. The switching circuit produced a optoisolators. The switching circuit produced a transient anode current (t approximately 15 microseconds) which was subtracted for measurement of recoveries of < 50-100 microseconds. The apparatus was coupled to an inverted microscope for measurement of fluorescence by epi-illumination or total internal reflection. Instrument performance was evaluated by measurement of the rapid fluorescence recoveries of fluorescein and fluorescein-dextrans in solutions and living cells.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Cells, Cultured
  • Fibroblasts
  • Fluoresceins / chemistry
  • Fluorescent Dyes / chemistry
  • Histocytochemistry / instrumentation*
  • Mice
  • Microscopy, Fluorescence
  • Software
  • Solutions
  • Spectrometry, Fluorescence / instrumentation*

Substances

  • Fluoresceins
  • Fluorescent Dyes
  • Solutions