The spiral Ames assay, an automated approach to bacterial mutagenicity testing which simplifies the test procedure and reduces the amount of drug required to generate mutagenic dose-response information, has been evaluated and validated for routine screening. The spiral plater delivers the Salmonella bacteria, exogenous metabolic activation system and drug to the surface of a rotating agar plate one on top of another in such a way that a uniform density of bacteria is exposed to a logarithmically decreasing volume of drug. Following an incubation of 48 hr at 37 degrees C, the plates are scanned by a laser counter, and the data are subjected to a computerized analysis. Petri plates of 15 cm diameter were used to provide a concentration range of about 250-fold per plate. The Salmonella were concentrated 20-fold to increase sensitivity. Thirty-eight compounds from a variety of chemical classes, including both pharmaceuticals and known mutagens of moderate to strong potency, were tested in both the spiral and the standard pour-plate assays. There was overall test agreement on positive or negative results for 82% of the compounds tested. When only the results from strains TA98 plus TA100 were considered, the agreement was 87%. When positive results were obtained, the fold increase over vehicle control was on average twice as great for the spiral assay compared to the pour-plate assay. It was concluded that the two assay procedures generally provided comparable results, with the spiral assay being somewhat more sensitive in terms of dose-response than the pour-plate assay.