Glucose oxidase electrodes: effect of hydrogen peroxide on enzyme activity?

Biosens Bioelectron. 1994;9(1):65-71. doi: 10.1016/0956-5663(94)80016-2.

Abstract

This paper presents a simple procedure to assess the amount of hydrogen peroxide arising in amperometric glucose oxidase sensors. At an external glucose concentration of 30 mmol/l, which represents the linear range of calibration, a hydrogen peroxide concentration of approximately 0.18 mmol/l is generated. Taking into consideration that in medical applications the main range of interest would be within euglycaemia which is between 4 and 9 mmol/l, it is concluded that in the type of glucose electrode investigated, hydrogen peroxide produced during enzymatic glucose measurement does not appreciably damage the enzyme glucose oxidase.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biosensing Techniques*
  • Glucose Oxidase*
  • Hydrogen Peroxide / chemistry*
  • Reproducibility of Results

Substances

  • Hydrogen Peroxide
  • Glucose Oxidase