Variations in power output of the Piezon-Master 400 ultrasonic endodontic unit

Int Endod J. 1994 Jan;27(1):26-31. doi: 10.1111/j.1365-2591.1994.tb00225.x.

Abstract

The present study was undertaken to see if there was any variability in the power output of Piezon-Master 400 ultrasonic files when driven using different generators, tranducers and file holders. The displacement amplitude of the oscillating tip of the file in air was used as a measure of the power output. The results showed that there was considerable variability in the power output of Piezon-Master 400 ultrasonic files of similar size and length when driven using different generators, transducers and file holders. In consideration of this, it is recommended that a calibration device be incorporated in the ultrasonic unit so that the operator will have some knowledge of when the unit is working at its maximum efficiency.