p-type silicon detector for brachytherapy dosimetry

Med Phys. 1995 Jun;22(6):835-9. doi: 10.1118/1.597486.

Abstract

The sensitivity of a cylindrical p-type silicon detector was studied by means of air and water measurements using different photon beams. A lead filter cap around the diode was used to minimize the dependence of the detector response as a function of the brachytherapy photon energy. The radial dose distribution of a high-activity 192Ir source in a brachytherapy phantom was measured by means of the shielded diode and the agreement of these data with theoretical evaluations confirms the method used to compensate diode response in the intermediate energy range. The diode sensitivity was constant over a wide range of dose rates of clinical interest; this allowed one to have a small detector calibrated in terms of absorbed dose in a medium. Theoretical evaluations showed that a single shielding filter around the p-type diode is sufficient to obtain accurate dosimetry for 192Ir, 137Cs, and 60Co brachytherapy sources.