The Effect of Different Intraoral Scanners on The Accuracy of Bite Registration in Edentulous Maxillary and Mandibular Arches

J Dent. 2024 May 10:105050. doi: 10.1016/j.jdent.2024.105050. Online ahead of print.

Abstract

Objectives: The objective of this study was to use in vitro models to examine the bite registration accuracy of four different intraoral scanners (IOS) for edentulous maxillary and mandibular arches. The objective was to assess the trueness and precision of the IOS and determine if there were significant differences between them.

Methods: An Asiga Max UV 3D printer was used to print maxillary and mandibular edentulous models based on the shape of Frasaco models (artificial dental arch models). Four dental implants were placed symmetrically in both models using Straumann BLT RC implants. Digital impressions were taken with Primescan, Trios 3, Trios 4, and Medit i500 intraoral scanners (n = 10 for each IOS). Digital bite registrations were made, and scanning data was exported in STL format. The accuracy of the interarch distance (the distance between the metrological spheres attached to the mandibular and maxillary models) was estimated for each IOS.

Results: The results showed significant differences in trueness and precision between different IOS (p <.05), except Medit i500 and Trios 3 (p >.05). Primescan provided the most accurate results, followed by Medit i500, Trios 3, and Trios 4, respectively.

Conclusions: within the limitations of this study, the IOS type affects the accuracy of interocclusal bite registration in in vitro design. Only Primescan achieved clinically acceptable accuracy for the interocclusal recording of edentulous arches.

Clinical relevance: The comparison of the accuracy of bite registration between different intraoral scanners will help increase the efficiency of the clinical application of digitalized interarch registration.

Keywords: Accuracy; Digital bite registration; Intraoral scanner; Precision; Prosthodontics; Trueness; edentulous.