Effect of particle size on liver MRI R 2 * relaxometry: Monte Carlo simulation and phantom studies

Magn Reson Med. 2024 May 9. doi: 10.1002/mrm.30154. Online ahead of print.

Abstract

Purpose: To investigate the effect of particle size on liver R 2 * $$ {\mathrm{R}}_2^{\ast } $$ by Monte Carlo simulation and phantom studies at both 1.5 T and 3.0 T.

Methods: Two kinds of particles (i.e., iron sphere and fat droplet) with varying sizes were considered separately in simulation and phantom studies. MRI signals were synthesized and analyzed for predicting R 2 * $$ {\mathrm{R}}_2^{\ast } $$ , based on simulations by incorporating virtual liver model, particle distribution, magnetic field generation, and proton movement into phase accrual. In the phantom study, iron-water and fat-water phantoms were constructed, and each phantom contained 15 separate vials with combinations of five particle concentrations and three particle sizes. R 2 * $$ {\mathrm{R}}_2^{\ast } $$ measurements in the phantom were made at both 1.5 T and 3.0 T. Finally, differences in R 2 * $$ {\mathrm{R}}_2^{\ast } $$ predictions or measurements were evaluated across varying particle sizes.

Results: In the simulation study, strong linear and positively correlated relationships were observed between R 2 * $$ {\mathrm{R}}_2^{\ast } $$ predictions and particle concentrations across varying particle sizes and magnetic field strengths ( r 0.988 $$ r\ge 0.988 $$ ). The relationships were affected by iron sphere size ( p < 0.001 $$ p<0.001 $$ ), where smaller iron sphere size yielded higher predicted R 2 * $$ {\mathrm{R}}_2^{\ast } $$ , whereas fat droplet size had no effect on R 2 * $$ {\mathrm{R}}_2^{\ast } $$ predictions ( p 0.617 $$ p\ge 0.617 $$ ) for constant total fat concentration. Similarly, the phantom study showed that R 2 * $$ {\mathrm{R}}_2^{\ast } $$ measurements were relatively sensitive to iron sphere size ( p 0.004 $$ p\le 0.004 $$ ) unlike fat droplet size ( p 0.223 $$ p\ge 0.223 $$ ).

Conclusion: Liver R 2 * $$ {\mathrm{R}}_2^{\ast } $$ is affected by iron sphere size, but is relatively unaffected by fat droplet size. These findings may lead to an improved understanding of the underlying mechanisms of R 2 * $$ {\mathrm{R}}_2^{\ast } $$ relaxometry in vivo, and enable improved quantitative MRI phantom design.

Keywords: Monte Carlo simulations; fat droplet; iron; liver R2*; particle size; phantom.