Assessing changes in regional cerebral hemodynamics in adults with a high-density full-head coverage time-resolved near-infrared spectroscopy device

J Biomed Opt. 2024 Jun;29(Suppl 3):S33302. doi: 10.1117/1.JBO.29.S3.S33302. Epub 2024 May 3.

Abstract

Significance: Cerebral oximeters have the potential to detect abnormal cerebral blood oxygenation to allow for early intervention. However, current commercial systems have two major limitations: (1) spatial coverage of only the frontal region, assuming that surgery-related hemodynamic effects are global and (2) susceptibility to extracerebral signal contamination inherent to continuous-wave near-infrared spectroscopy (NIRS).

Aim: This work aimed to assess the feasibility of a high-density, time-resolved (tr) NIRS device (Kernel Flow) to monitor regional oxygenation changes across the cerebral cortex during surgery.

Approach: The Flow system was assessed using two protocols. First, digital carotid compression was applied to healthy volunteers to cause a rapid oxygenation decrease across the ipsilateral hemisphere without affecting the contralateral side. Next, the system was used on patients undergoing shoulder surgery to provide continuous monitoring of cerebral oxygenation. In both protocols, the improved depth sensitivity of trNIRS was investigated by applying moment analysis. A dynamic wavelet filtering approach was also developed to remove observed temperature-induced signal drifts.

Results: In the first protocol (28±5 years; five females, five males), hair significantly impacted regional sensitivity; however, the enhanced depth sensitivity of trNIRS was able to separate brain and scalp responses in the frontal region. Regional sensitivity was improved in the clinical study given the age-related reduction in hair density of the patients (65±15 years; 14 females, 13 males). In five patients who received phenylephrine to treat hypotension, different scalp and brain oxygenation responses were apparent, although no regional differences were observed.

Conclusions: The Kernel Flow has promise as an intraoperative neuromonitoring device. Although regional sensitivity was affected by hair color and density, enhanced depth sensitivity of trNIRS was able to resolve differences in scalp and brain oxygenation responses in both protocols.

Keywords: carotid compression; cerebral oximetry; high-density near-infrared spectroscopy; near-infrared spectroscopy; phenylephrine; time-resolved near-infrared spectroscopy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Brain / blood supply
  • Brain / diagnostic imaging
  • Cerebrovascular Circulation* / physiology
  • Equipment Design
  • Female
  • Hemodynamics / physiology
  • Humans
  • Male
  • Oximetry / instrumentation
  • Oximetry / methods
  • Oxygen / blood
  • Oxygen / metabolism
  • Spectroscopy, Near-Infrared* / instrumentation
  • Spectroscopy, Near-Infrared* / methods

Substances

  • Oxygen