Comparison of B-Scan ultrasonography, ultra-widefield fundus imaging, and indirect ophthalmoscopy in detecting retinal breaks in cataractous eyes

Eye (Lond). 2024 Apr 24. doi: 10.1038/s41433-024-03093-2. Online ahead of print.

Abstract

Background/objectives: To evaluate the diagnostic performance of B-scan kinetic ultrasonography (USG), standard ultra-widefield (UWF) imaging, and indirect ophthalmoscopy (IDO) in retinal break detection in cataractous eyes.

Subjects/methods: We consecutively enrolled 126 cataract patients (including 246 eyes) with no comorbidities that could decrease best corrected visual acuity (BCVA). Three index tests (USG, nonmydriatic UWF, and mydriatic IDO) were performed preoperatively to screen for retinal breaks. One week after cataract extraction, a dilated IDO examination was repeated for the definitive diagnosis of retinal break as the reference standard. The sensitivity, specificity, Youden index (YI), and predictive values of each index test were calculated according to postoperative ophthalmoscopic findings. A deep-learning nomogram was developed to quantify the risk of retinal break presence using patients' baseline data and findings reported from preoperative ophthalmic tests.

Results: Fifty-two eyes (21%) were excluded from appropriate preoperative UWF imaging because of massive lens opacity. The BCVA cutoff point with maximum YI indicating UWF applicability was 0.6 logMAR (YI = 0.3; area under curve [AUC] = 0.7). Among all 246 eyes, preoperative IDO, USG, and UWF showed fair interobserver agreement (all κ > 0.2). According to postoperative IDO findings, the index tests with the highest sensitivity and specificity were USG (100%) and preoperative IDO (99%), respectively.

Conclusions: For cataractous eyes without vision-impairing comorbidities, a BCVA better than 0.6 logMAR (Snellen acuity, 20/80) allows for appropriate nonmydriatic standard UWF imaging. In a high-volume clinic equipped with skilled ophthalmic examiners, screening with USG followed by directed IDO allows the efficient identification of retinal breaks in cataractous eyes.