Performance of a Norfentanyl Immunoassay in Specimens with Low Concentrations of Fentanyl and/or Norfentanyl

J Appl Lab Med. 2024 Apr 24:jfae036. doi: 10.1093/jalm/jfae036. Online ahead of print.

Abstract

Background: Many fentanyl immunoassays are limited in their ability to detect norfentanyl. Urine specimens collected from individuals who have been exposed to fentanyl frequently have detectable concentrations of norfentanyl (≥2 ng/mL) but low concentrations of fentanyl (<2 ng/mL) by LC-MS/MS. The Lin-Zhi Fentanyl II Immunoassay (Lin-Zhi) claims 100% cross-reactivity with norfentanyl and therefore may detect exposure missed by other assays.

Methods: In addition to verifying the manufacturer's analytical sensitivity claims, we selected 92 urine specimens with low-positive Lin-Zhi results (1-99 absorbance units, lowest 10%) for analysis by the Immunalysis Health Equity Impact Assessment and ARK II fentanyl methods. The accuracy of the 3 immunoassays was compared to LC-MS/MS as the reference method.

Results: Spiking studies using purified fentanyl and norfentanyl and a set of 100 consecutive specimens confirmed the manufacturer's claims of limit of detection for fentanyl (3.8 ng/mL) and norfentanyl (5.0 ng/mL). However, the 92 low-positive patient specimens demonstrated concentrations of norfentanyl and fentanyl below 2.0 ng/mL by LC-MS/MS, with 47 (51%) having only norfentanyl detected. When comparing Lin-Zhi to the Immunalysis and ARK II immunoassays, only 27 (29%) of the 92 specimens were concordant. Fifty-two (57%) of the specimens were positive by LC-MS/MS and Lin-Zhi but false negative by one or both other immunoassays. Seven specimens (8%) were positive by Lin-Zhi but negative by the other immunoassays and had undetectable concentrations (<2 ng/mL) of fentanyl and norfentanyl by LC-MS/MS.

Conclusions: The clinical sensitivity of the Lin-Zhi exceeds the manufacturer's claims, providing results comparable to LC-MS/MS methods.