A novel surgical technique for removal of intramedullary nails when there is a failure of the proximal extraction device: Report of two cases and review of literature

Injury. 2024 Jun;55(6):111533. doi: 10.1016/j.injury.2024.111533. Epub 2024 Apr 4.

Abstract

Diaphyseal fractures of the tibia is a common injury with an incidence of 16.9/100,000 population. Surgical treatment with intramedullary interlocking nail is the treatment of choice in closed diaphyseal tibial injuries. Removal of the nail is required in non-union, infected implant, implant failure or during exchange nailing procedure. Implant removal surgery presents some unique challenges to the Orthopaedic surgeons like broken nails, broken screws and failure of the extraction devices to engage with the threads of the nail due to stripping of the threads in the proximal part of the nail. Various techniques have been described by different authors to extract a broken distal piece of the nail. There are only a few articles described for the removal of an intramedullary nail when there is failure of the proximal extraction device. The failure of the proximal extraction device occurs due to the use of ill-fitting extraction bolts, cross threading and the presence of a very tight nail leading to stripping of the proximal threads. We describe a simple and novel technique with the use of the Stainless-Steel wire along with the extraction bolt when there is a failure of proximal extraction device for the extraction of cannulated or solid intramedullary nails.

Keywords: Failure of nail removal; Implant removal; Intramedullary nail; Proximal extraction device; Stainless-steel wire.

Publication types

  • Case Reports
  • Review

MeSH terms

  • Adult
  • Bone Nails*
  • Device Removal* / methods
  • Equipment Failure
  • Fracture Fixation, Intramedullary* / instrumentation
  • Fracture Fixation, Intramedullary* / methods
  • Humans
  • Male
  • Middle Aged
  • Tibial Fractures* / surgery