The Electrified Wire Technique in Complex Aortic Interventions: A Case Series

J Endovasc Ther. 2024 Apr 10:15266028241245341. doi: 10.1177/15266028241245341. Online ahead of print.

Abstract

Objectives: Electrosurgery has been long used in endovascular procedures, with only case reports in the aortic field. Our aim is to present a case series with the use of an electrified wire to perform catheter-based electrosurgery by applying external current through an electrocautery pen.

Methods: Single-center retrospective case series of all patients undergoing complex aortic surgery from October 2020 to August 2023, in whom the electrified wire technique was used: (1) Perforation of a dissection flap or left subclavian artery (LSA) in situ endograft fenestration-a 0.014" polytetrafluoroethylene (PTFE) insulated guidewire is detached from the insulation with a scalpel at the end and a cautery pen is here attached with a clamp. A curved tip catheter or sheath is positioned against the aortic flap or the endograft (through a left brachial access in this case) and the wire pushed, crossing the flap by activating the electrocautery pen and (2) slicing a dissection flap ("powered cheese-wire technique")-after same preparation as above, the middle section of the 0.014 guidewire is removed from the PTFE and bent into a V-shape. Once in the aorta, the guidewire crosses from the true lumen (TL) to the false lumen (FL) and a through-and-through access is obtained. Sheaths are positioned against the flap from both sides and moved up or down while the electricity is activated, slicing the flap and communicating both lumens. Technical success and technical-related complications were evaluated.

Results: Eleven cases concerning aortic dissections and 1 case of aortic atresia were treated. Four patients presented urgently, whereas the rest were planned procedures. Seven cases underwent perforation of a dissection flap, 2 cases underwent the powered cheese-wire technique, in 2 cases for an LSA in situ fenestration, and in 1 case to cross an aortic atresia at the aortic isthmus. The technique was in all cases successfully applied. No complications related to the technique occurred.

Conclusions: The "electrified wire" technique is a feasible and ready-available tool that can be safely used in complex aortic interventions, especially to perforate aortic tissue like dissection flaps or to perform in situ fenestrated repairs by perforation of the endograft fabric.

Clinical impact: The electrified wire technique described herein is a straightforward technique that uses readily available tools to perform electrosurgery. We present its use in complex aortic procedures. However, it could be envisioned for any vascular procedure that requires crossing of the vessel or even prosthetic material. As we have described in this series, when used along with an adequate properative planning, it can be a safe tool of great utility, as has already been demonstarted in the field of the interventional cardiology.

Keywords: aortic atresia; aortic dissection; dissection flap; electrosurgery; in situ fenestration.