Evaluation of cycloplegic and noncycloplegic performance of spot vision screener in detection of amblyopia risk factors using 2021 AAPOS guidelines

Photodiagnosis Photodyn Ther. 2024 Apr 2:46:104073. doi: 10.1016/j.pdpdt.2024.104073. Online ahead of print.

Abstract

Purpose: This research evaluates the effectiveness of the Spot Vision Screener (SVS) before and after cycloplegia to detect amblyogenic refractive errors in children.

Methods: Children ages 3 to 10 years old were screened by the SVS before and after cycloplegia. Sensitivity, specificity, positive and negative predictive value, paired t-test, Bland-Altman plot and receiver operating characteristic area under the curve were evaluated by comparing the results of the SVS (v3.0.05) measurements with the results of the cycloplegic Topcon autorefractometer according to the 2021 guidelines of the American Association for Pediatric Ophthalmology and Strabismus.

Results: Both eyes of 211 patients aged 3 to 10 years old were included. Regarding the amblyopia risk factors, the noncycloplegic SVS had 65.7 % sensitivity, 94.9 % specificity, 81.2 % positive predictive value and 89.3 % negative predictive value. The SVS's sensitivity increased from 65.7 % to 81.9 % with cycloplegia compared to noncycloplegic SVS results. The sensitivity detection of hyperopia was improved from 4.2 % to 100 % after cycloplegia. Areas under the receiver operator characteristic curve for noncycloplegic SVS and cycloplegic SVS were 0.506 (95 % CI, 0.395 to 0.646, p = 0737) and 0.905 (95 % CI, 0.915 to 0.971, p < 0.001) for hyperopia, respectively. Using the +1.64 D revised cutoff criteria for hyperopia increased sensitivity from 4.2 % to 78 %.

Conclusion: Noncycloplegic SVS measurements showed relatively high specificity in detecting amblyopia risk factors. The fact that noncycloplegic measurements have a very low sensitivity for hyperopia is an important weakness of the SVS, especially because hyperopia is the most frequently encountered refractive error in very young children. It should be noted that amblyogenic hyperopia may be overlooked by an SVS without cycloplegia.

Keywords: Amblyopia risk factors; Children; Cycloplegia; Refractive error; Spot vision screener.