Study of a plastic scintillating plate-based quality assurance system for pencil beam scanning proton beams

J Cancer Res Ther. 2024 Jan 1;20(1):85-92. doi: 10.4103/jcrt.jcrt_1344_22. Epub 2023 Apr 7.

Abstract

Introduction: The purpose of this study was to evaluate a plastic scintillating plate-based beam monitoring system to perform quality assurance (QA) measurements in pencil beam scanning proton beam.

Methods: Single spots and scanned fields were measured with the high-resolution dosimetry system, consisting of a plastic scintillation plate coupled to a camera in a dark box at the isocenter. The measurements were taken at 110-190 MeV beam energies with 30° gantry angle intervals at each energy. Spot positions were determined using the plastic scintillating plate-based dosimetry system at the isocenter for 70-230 MeV beam energies with 30° gantry angle intervals. The effect of gantry angle on dose distribution was also assessed by determining the scanning pattern for daily QA and 25 fields treated with intensity-modulated proton therapy.

Results: Spot size, field flatness, and field symmetry of plastic scintillating plate-based dosimetry system were consistent with EBT3 at all investigated energies and angles. In all investigated energies and angles, the spot size measured was ±10% of the average size of each energy, the spot position measured was within ±2 mm, field flatness was within ±2%, and field symmetry was within ±1%. The mean gamma passing rates with the 3%/3 mm gamma criterion of the scanning pattern and 25 fields were 99.2% and 99.8%, respectively.

Conclusions: This system can be effective for QA determinations of spot size, spot position, field flatness, and field symmetry over 360° of gantry rotation in a time- and cost-effective manner, with spatial resolution comparable to that of EBT3 film.

MeSH terms

  • Humans
  • Proton Therapy*
  • Protons
  • Radiometry
  • Radiotherapy Dosage

Substances

  • Protons