Dual-Frequency Impedance Matching Network Design Using Genetic Algorithm for Power Ultrasound Transducer

Micromachines (Basel). 2024 Feb 29;15(3):344. doi: 10.3390/mi15030344.

Abstract

Dual-frequency ultrasounds have demonstrated significant potential in augmenting thermal ablation efficiency for tumor treatment. Ensuring proper impedance matching between the dual-frequency transducer and the power amplifier system is imperative for equipment safety. This paper introduces a novel dual-frequency impedance matching network utilizing L-shaped topology and employing a genetic algorithm to compute component values. Implementation involved an adjustable capacitor and inductor network to achieve dual-frequency matching. Subsequently, the acoustic parameters of the dual-frequency HIFU transducer were evaluated before and after matching, and the effects of ultrasound thermal ablation with and without matching were compared. The proposed dual-frequency impedance matching system effectively reduced the standing wave ratio at the two resonance points while enhancing transmission efficiency. Thermal ablation experiments with matching circuits showed improved temperature rise efficiencies at both frequencies, resulting in an expanded ablation zone. The dual-frequency impedance matching method significantly enhances the transmission efficiency of the dual-frequency ultrasound system at two operational frequencies, thereby ensuring equipment safety. It holds promising prospects for application in dual-frequency ultrasound treatment.

Keywords: dual-frequency ultrasound; electrical impedance matching; genetic algorithm; power ultrasound transducer; ultrasound thermal ablation.