Effect of different framework materials on stresses induced in distal extension removable partial dentures: A strain gauge analysis

J Prosthodont. 2024 Mar 23. doi: 10.1111/jopr.13847. Online ahead of print.

Abstract

Purpose: Evaluation of the strain transmitted to the abutments and residual ridge by polyetherketoneketone material compared to the cobalt-chromium one in distal extension removable partial dentures (RPDs) to fulfill the objective of preservation of the supporting structures.

Materials and methods: A virtual model simulating a Kennedy class I partially edentulous mandibular arch was designed. Two models, one for each group, were printed. Five RPDs were made in each group. In group CR, the framework was milled from a cobalt-chromium alloy. While in group PK, it was milled from a polyetherketoneketone blank. Strain gauge rosettes were bonded distal to the last abutment and posteriorly in the distal end of the residual ridge. Unilateral vertical and oblique loadings were applied. Mann-Whitney U test was used for inter-group comparisons while the Friedman test was used for intra-group comparisons and corrected by Wilcoxon Signed-Rank Sum. The significance level was set at p ≤ 0.05.

Results: During unilateral vertical load application, a statistically significant difference was found between both groups distal to the abutment in the loaded and unloaded sides as well as the residual ridge on the unloaded side. During oblique load application, a statistically significant difference was found between both groups in all slots.

Conclusion: Polyetherketoneketone material induces less stress on the abutments and more stress on the residual ridges compared to the cobalt-chromium ones. Therefore, it may be recommended for weak abutments supporting RPDs.

Keywords: Co‐Cr; PEKK; Polyetherketoneketone; cobalt‐chromium; framework; milling; stress analysis.