Insulin biotrapping using plasmofluidic optical fiber chips: A benchmark

Biosens Bioelectron. 2024 Jun 15:254:116189. doi: 10.1016/j.bios.2024.116189. Epub 2024 Mar 12.

Abstract

Plasmonic optical fiber-based biosensors are currently in their early stages of development as practical and integrated devices, gradually making their way towards the market. While the majority of these biosensors operate using white light and multimode optical fibers (OFs), our approach centers on single-mode OFs coupled with tilted fiber Bragg gratings (TFBGs) in the near-infrared wavelength range. Our objective is to enhance surface sensitivity and broaden sensing capabilities of OF-based sensors to develop in situ sensing with remote interrogation. In this study, we comprehensively assess their performance in comparison to the gold-standard plasmonic reference, a commercial device based on the Kretschmann-Raether prism configuration. We present their refractive index sensitivity and their capability for insulin sensing using a dedicated microfluidics approach. By optimizing a consistent surface biotrapping methodology, we elucidate the dynamic facets of both technologies and highlight their remarkable sensitivity to variations in bulk and surface properties. The one-to-one comparison between both technologies demonstrates the reliability of optical fiber-based measurements, showcasing similar experimental trends obtained with both the prismatic configuration and gold-coated TFBGs, with an even enhanced limit of detection for the latter. This study lays the foundation for the detection of punctual molecular interactions and opens the way towards the detection of spatially and temporally localized events on the surface of optical probes.

Keywords: Biacore; Biosensor; Insulin detection; Optical fiber; Plasmonics; Tilted fiber bragg grating (TFBG).

MeSH terms

  • Benchmarking
  • Biosensing Techniques* / methods
  • Insulin
  • Optical Fibers*
  • Reproducibility of Results

Substances

  • Insulin