Pre-excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre-ENCODE)

Magn Reson Med. 2024 Mar 19. doi: 10.1002/mrm.30068. Online ahead of print.

Abstract

Purpose: To evaluate the use of pre-excitation gradients for eddy current-nulled convex optimized diffusion encoding (Pre-ENCODE) to mitigate eddy current-induced image distortions in diffusion-weighted MRI (DWI).

Methods: DWI sequences using monopolar (MONO), ENCODE, and Pre-ENCODE were evaluated in terms of the minimum achievable echo time (TE min $$ {}_{\mathrm{min}} $$ ) and eddy current-induced image distortions using simulations, phantom experiments, and in vivo DWI in volunteers ( N = 6 $$ N=6 $$ ).

Results: Pre-ENCODE provided a shorter TE min $$ {}_{\mathrm{min}} $$ than MONO (71.0 ± $$ \pm $$ 17.7ms vs. 77.6 ± $$ \pm $$ 22.9ms) and ENCODE (71.0 ± $$ \pm $$ 17.7ms vs. 86.2 ± $$ \pm $$ 14.2ms) in 100 % $$ \% $$ of the simulated cases for a commercial 3T MRI system with b-values ranging from 500 to 3000 s/mm 2 $$ {}^2 $$ and in-plane spatial resolutions ranging from 1.0 to 3.0mm 2 $$ {}^2 $$ . Image distortion was estimated by intravoxel signal variance between diffusion encoding directions near the phantom edges and was significantly lower with Pre-ENCODE than with MONO (10.1 % $$ \% $$ vs. 22.7 % $$ \% $$ , p = 6 - 5 $$ p={6}^{-5} $$ ) and comparable to ENCODE (10.1 % $$ \% $$ vs. 10.4 % $$ \% $$ , p = 0 . 12 $$ p=0.12 $$ ). In vivo measurements of apparent diffusion coefficients were similar in global brain pixels (0.37 [0.28,1.45] × 1 0 - 3 $$ \times 1{0}^{-3} $$ mm 2 $$ {}^2 $$ /s vs. 0.38 [0.28,1.45] × 1 0 - 3 $$ \times 1{0}^{-3} $$ mm 2 $$ {}^2 $$ /s, p = 0 . 25 $$ p=0.25 $$ ) and increased in edge brain pixels (0.80 [0.17,1.49] × 1 0 - 3 $$ \times 1{0}^{-3} $$ mm 2 $$ {}^2 $$ /s vs. 0.70 [0.18,1.48] × 1 0 - 3 $$ \times 1{0}^{-3} $$ mm 2 $$ {}^2 $$ /s, p = 0 . 02 $$ p=0.02 $$ ) for MONO compared to Pre-ENCODE.

Conclusion: Pre-ENCODE mitigated eddy current-induced image distortions for diffusion imaging with a shorter TE min $$ {}_{\mathrm{min}} $$ than MONO and ENCODE.

Keywords: diffusion; eddy currents; time-optimal.