Adaptive estimation of continuous gait phase based on capacitive sensors

Wearable Technol. 2022 Jun 17:3:e11. doi: 10.1017/wtc.2022.4. eCollection 2022.

Abstract

Continuous gait phase plays an important role in robotic prosthesis control. In this paper, we have conducted the offline adaptive estimation (at different speeds and on different ramps) of continuous gait phase of robotic transtibial prosthesis based on the adaptive oscillators. We have used the capacitive sensing method to record the deformation of the muscles. Two transtibial amputees joined in this study. Based on the strain signals of the prosthetic foot and the capacitive signals of the residual limb, the maximum and minimum of estimation errors are 0.80 rad and 0.054 rad, respectively, and their corresponding ratios in one gait cycle are 1.27% and 0.86%, respectively. This paper proposes an effective method to estimate the continuous gait phase based on the capacitive signals of the residual muscles, which provides a basis for the continuous control of robotic transtibial prosthesis.

Keywords: capacitive sensors; continuous gait phase; robotic transtibial prosthesis.