Determination of Critical Dimensions of Microchannels to Ensure the Electrokinetic Biomolecule Preconcentration: Analytical and Numerical Studies

Langmuir. 2024 Mar 19;40(11):6051-6064. doi: 10.1021/acs.langmuir.4c00300. Epub 2024 Mar 4.

Abstract

Preconcentration of biomolecules based on ion concentration polarization (ICP) has been splendidly applied to various biomedical and chemical processes. However, in many circumstances, biomolecule preconcentration could not occur due to the lack of full studies on the preconcentration mechanism, especially on the effect of microchannel dimensions. In this work, we provide analytical studies on the critical dimensions (minimum and maximum) of microchannels for the preconcentration of biomolecules. These formulas are verified with the numerical results by fully solving the coupled governing equations: Poisson-Nernst-Planck and Navier-Stokes experiments with appropriate boundary conditions and assumptions. In addition, we examine the impact of operational parameters, such as electric potentials and critical external pressures, on the formation of the preconcentration. Moreover, two important results are provided for the first time, including the position of the preconcentrated biomolecule region and the concentration enhancement factor of the biomolecules. These analytical and numerical results are consistent with experimental observations and, therefore, could provide sharp insight into the mechanism of biomolecule preconcentration and give useful guidelines to better design and optimize ICP-based microfluidic preconcentration devices.

MeSH terms

  • Lab-On-A-Chip Devices*