A novel intraocular pressure predicting method based on hyperelastic mechanical model of cornea

J Mech Behav Biomed Mater. 2024 May:153:106475. doi: 10.1016/j.jmbbm.2024.106475. Epub 2024 Feb 26.

Abstract

Measuring intraocular pressure (IOP) is crucial and remains challenging in diagnosing glaucoma, as it is associated with cornea deformation during inflation. In this study, a three-dimensional analytical model based on hyperelastic constitutive relationship to predict correlation between cornea vertex displacement and the IOP is proposed. The analytical model is validated by rigorous experiments. Rabbit corneas were selected for this study and their mechanical properties were obtained using uniaxial tensile tests. To mimic the environment in which the cornea exists, an artificial anterior chamber equipped with water-injection pipelines was constructed to study the relationship between the corneal vertex displacement with IOP value in practical situation. The experimental results of rabbits corneas prove that the IOP can be deduced based on the measured corneal vertex displacement by the analytical model. Furthermore, subtle difference occurs when comparing the calculated human IOPs with those measured by medical equipment, demonstrating that the proposed method is suitable for monitoring the IOP of human. This novel IOP predicting method provides new inspiration for the design of eyepieces, as well as the preoperative preparation for laser surgery and evaluation of corneal damage.

Keywords: Cornea deformation; Hyperelastic model; Inflating test; Intraocular pressure.

MeSH terms

  • Animals
  • Cornea
  • Corneal Injuries*
  • Humans
  • Intraocular Pressure*
  • Rabbits
  • Tonometry, Ocular / methods