Reproducibility of consecutive automated telemetric noctodiurnal IOP profiles as determined by an intraocular implant

Br J Ophthalmol. 2024 Feb 26:bjo-2022-323080. doi: 10.1136/bjo-2022-323080. Online ahead of print.

Abstract

Background: Intraocular pressure (IOP) monitoring in glaucoma management is evolving with novel devices. We investigated the reproducibility of 24 hour profiles on two consecutive days and after 30 days of self-measurements via telemetric IOP monitoring.

Methods: Seven primary patients with open-angle glaucoma previously implanted with a telemetric IOP sensor in one eye underwent automatic measurements throughout 24 hours on two consecutive days ('day 1' and 'day 2'). Patients wore an antenna adjacent to the study eye connected to a reader device to record IOP every 5 min. Also, self-measurements in six of seven patients were collected for a period of 30 days. Analysis included calculation of hourly averages to correlate time-pairs of day 1 versus day 2 and the self-measurements vers day 2.

Results: The number of IOP measurements per patient ranged between 151 and 268 on day 1, 175 and 268 on day 2 and 19 and 1236 during 30 days of self-measurements. IOP time-pairs of automatic measurements on day 1 and day 2 were significantly correlated at the group level (R=0.83, p<0.001) and in four individual patients (1, 2, 6 and 7). IOP time-pairs of self-measurements and day 2 were significantly correlated at the group level (R=0.4, p<0.001) and in four individual patients (2, 5, 6 and 7).

Conclusions: Twenty-four hour automatic measurements of IOP are correlated on consecutive days and, though to a lesser degree, with self-measurements. Therefore a virtual 24-hour IOP curve might be constructed from self-measurements. Both options provide an alternative to frequent in-office IOP measurements.

Keywords: glaucoma; intraocular pressure; telemedicine.