Impact of scanning range and image count on the precision of digitally recorded intermaxillary relationships in interocclusal record using intraoral scanner

J Oral Sci. 2024 Apr 16;66(2):111-115. doi: 10.2334/josnusd.23-0379. Epub 2024 Feb 23.

Abstract

Purpose: The effect of scan range and the number of scanned images on the precision of in vivo intermaxillary relationship reproduction was evaluated using digital scans acquired with an intraoral scanner.

Methods: The study involved 15 participants with normal occlusion. Two different interocclusal recording settings were employed using the intraoral scanner (TRIOS 4): 'MIN,' focusing on the minimal scan range of the first molar region, and 'MAX,' including the scan range from the right first premolar to the right second molar. These settings were combined with three different image counts, resulting in six experimental conditions. Interocclusal recordings were performed four times for each condition. Dimensional discrepancies between datasets were analyzed using three-dimensional morphometric software and compared using two-way analysis of variance.

Results: Median dimensional discrepancies (interquartile range; IQR) of 39.2 (30.7-49.4), 42.2 (32.6-49.3), 30.3 (26.8-44.1), 20.1 (16.0-34.8), 21.8 (19.0-25.1), and 26.6 (19.9-34.5) µm were found for MIN/200, MIN/400, MIN/600, MAX/200, MAX/400, and MAX/600, respectively. Significant differences in dimensional discrepancies according to scan range were found. Wilcoxon signed-rank test showed significant differences between MAX and MIN (P < 0.01).

Conclusion: Scan range may affect the precision of intermaxillary relationship reproduction. Thus, scanning of the most extensive region practically achievable is recommended.

Keywords: digital scan; interocclusal record; intraoral scanner; precision.