Comparison of Mechanical Behavior of Clear Aligner and Rapid Palatal Expander on Transverse Plane: An In Vitro Study

Bioengineering (Basel). 2024 Jan 23;11(2):103. doi: 10.3390/bioengineering11020103.

Abstract

(1) Background: This study aims to investigate, within a controlled laboratory environment, the magnitude of the transversal load and the force decay over time produced by clear aligners in comparison to a Rapid Palatal Expander (RPE). (2) Methods: Resin models of a dental maxillary arch, additively manufactured from an intraoral scan, were inserted in a testing machine with uniaxial load cells to measure the force trend over time expressed by RPE and clear aligners. The mechanical load was recorded during a certain timeframe for both appliances. (3) Results: The force expressed by the RPE ranged from 30 to 50 N for each activation, decreasing with a nonlinear pattern over time. The force expressed by the clear aligner ranged from 3 to 5 N, decreasing with a linear pattern over time. In contrast, the force generated by the clear aligner fell within the range of 3 to 5 N, showing a linear reduction in force magnitude over the observed period of time. (4) Conclusions: The RPE exerted a force magnitude approximately ten times greater than that generated by clear aligners. Nevertheless, it is essential to acknowledge that the oral environment can significantly influence these results. These limitations underscore the need for caution when applying these findings to clinical settings.

Keywords: clear aligner; malocclusion; maxillary expansion.