Influence of the physical pressure of an ultrasound probe on shear-wave elastography measurements of the gastrocnemius muscle in a paediatric population: a non-interventional cohort study

Pol J Radiol. 2024 Jan 16:89:e24-e29. doi: 10.5114/pjr.2024.134437. eCollection 2024.

Abstract

Purpose: The aim of our study is to evaluate the impact of ultrasound probe mechanical pressure on the stiffness of the gastrocnemius muscle in a healthy paediatric population. As far as we know, there has been no previous qualitative in vivo study on the impact of probe pressure on muscle shear-wave elastography results with objective evaluation of compression in the paediatric population.

Material and methods: In this cohort study, a group of 22 children (mean age 8.99 years, SD 2.74, 11 males) underwent elastography of the gastrocnemius muscle of the dominant leg. A custom-made, 3-dimensional printed probe cover was used to measure the mechanical pressure of the probe on tissues.

Results: The obtained results were related to the age, sex, BMI, and calf circumference of the subjects. We observed a significant difference in the stiffness parameter at a pressure of 1 N, with a further increase if force was increased (p < 0.001). A significant, very weak positive correlation of age and stiffness was observed (p < 0.001, r2 = 0.022). There was no significant correlation of stiffness, BMI, and calf circumference.

Conclusions: The use of compression during muscle elastography in children causes a significant bias in results, regardless of age, sex, BMI, or calf size.

Keywords: elastography; muscles; paediatrics; shear wave elastography; ultrasound.