When bias becomes part of imprecision: how to use analytical performance specifications to determine acceptability of lot-lot variation and other sources of possibly unacceptable bias

Clin Chem Lab Med. 2024 Feb 8. doi: 10.1515/cclm-2023-1303. Online ahead of print.

Abstract

ISO 15189 requires laboratories to estimate the uncertainty of their quantitative measurements and to maintain them within relevant performance specifications. Furthermore, it refers to ISO TS 20914 for instructions on how to estimate the uncertainty and what to take into consideration when communicating uncertainty of measurement with requesting clinicians. These instructions include the responsibility of laboratories to verify that bias is not larger than medically significant. If estimated to be larger than acceptable, such bias first needs to be eliminated or (temporarily) corrected for. In the latter case, the uncertainty of such correction becomes part of the estimation of the total measurement uncertainty. If small enough to be acceptable, bias becomes part of the long term within laboratory random variation. Sources of possible bias are (not limited to) changes in reagent or calibrator lot variation or calibration itself. In this paper we clarify how the rationale and mathematics from an EFLM WG ISO/A position paper on allowable between reagent lot variation can be applied to calculate whether bias can be accepted to become part of long-term imprecision. The central point of this rationale is to prevent the risk that requesting clinicians confuse changes in bias with changes in the steady state of their patients.