Subcutaneous injection of hyaluronic acid leading to emboliom and recanalization process monitored in real time by three-dimensional photoacoustic imaging

Heliyon. 2024 Jan 20;10(3):e25066. doi: 10.1016/j.heliyon.2024.e25066. eCollection 2024 Feb 15.

Abstract

This study describes a method for real-time examination of the microvascular system based on the three-dimensional photoacoustic imaging system to prevent arterial complications, especially vascular embolism, during hyaluronic acid (HA) injections. Chicken embryos were used to simulate the superficial blood vessels of human skin, and then the target area was imaged by the photoacoustic imaging system for three-dimensional vascular imaging, and then the syringe and blood vessels were monitored, and the syringe angle and penetration depth were adjusted in time using an injection device to avoid puncturing the arterial vasculature and clogging the blood vessels. HA was then injected into smaller vessels on the dorsum of the tongue in mice and into thicker vessels on the dorsal portion of the tongue in rats to mimic embolization, and the post-operative recovery was reflected by the changes in the pixel dots of the extracted part of the blocked blood vessels, and it was observed that the blood flow in the area of the fine vessels was restored in about 3 days, whereas blood flow in the area of the large vessels was restored in only about 1 h. The method presented in this paper allows precise guidance of injectable filler HA, which has good application prospects in improving the safety of injection micro-plastic surgery and reducing the experience requirements for medical personnel.

Keywords: Filler injection; Hyaluronic acid; Photoacoustic imaging; Vascular compromise.