Background: Post-myocardial infarction ventricular septal rupture (VSR) is a rare and severe complication of myocardial infarction. To find early mortality (<30 days) risk factors of device VSR closure and to evaluate its medium-term outcome.
Methods: Multicenter retrospective analysis on all 46 consecutive patients with percutaneous (n = 43) or hybrid (n = 3) VSR closure in 2000-2020 with various nitinol wire mesh occluders. Medical records, hemodynamic data, procedure results, short- and mid-term follow-up were analyzed (4.8 ± 3.7 years, range: 0.1-15, available in 61.7% of patients). Of the patients, 34.8% underwent VSR closure in acute phase (<21 days after VSR occurrence), 17.4% underwent device closure due to significant residual shunt after previous VSR surgery.
Results: Success rate was 78.3%. More than moderate residual shunt, major complications, and early surgical reintervention affected 18.9%, 15.2% (including 2 intra-procedural deaths), and 21.7% of patients, respectively. Early mortality was 26.1% (13.9% in successful vs. 70% in unsuccessful closure; p < 0.001). Older age, need for intra-aortic balloon counterpulsation, severe complications, and procedural failure were identified as risk factors for early mortality. Among patients who survived the early period, the 5-year survival rate was 57.1%. NYHA class improved in 88.2% patients at the latest follow-up.
Conclusions: Procedure of VSR device closure demonstrates an acceptable technical success rate; however, the incidence of severe complications and early mortality is notably high. Older patients in poor hemodynamic condition and those with unsuccessful occluder deployment are particularly at a higher risk of a fatal outcome. The prognosis after early survival is promising.
Keywords: Early mortality; Follow-up; Hybrid closure; Percutaneous closure; Post-myocardial infarction ventricular septal rupture.
Copyright © 2024 Elsevier Ireland Ltd. All rights reserved.