A Comparison of Functional Luminal Imaging Probe With High-resolution Manometry, Timed Barium Esophagram, and pH Impedance Testing to Evaluate Functional Luminal Imaging Probe's Diagnostic Capabilities

J Clin Gastroenterol. 2024 Jan 15. doi: 10.1097/MCG.0000000000001966. Online ahead of print.

Abstract

Objective: The aims of this study are to determine the functional luminal imaging probe's (FLIP) diagnostic utility by comparing FLIP measurements with results from other esophageal evaluation standards.

Background: The FLIP is an esophageal evaluation technique performed at the time of endoscopy. Few studies have evaluated FLIP diagnostic capabilities compared with the established testing techniques, including high-resolution manometry (HRIM), time barium esophagram (TBE), and 24-hour impedance-pH monitoring.

Patients and methods: A retrospective review was performed for 413 preintervention patients who underwent FLIP testing during endoscopy. Data from HRIM, 24-hour pH monitoring, and TBE were compared.

Results: Abnormal Distensibility Index (DI) was associated with abnormal integrated relaxation pressure (IRP; P = 0.003). Average DI was higher in patients with abnormal IRP (>15 mm Hg) when a hiatal hernia was present (P = 0.025). The total agreement between correlated diagnoses from FLIP and HRIM was 33.5%. DI was not associated with acid exposure time on pH monitoring. Agreement between FLIP and TBE was 49% with a sensitivity of 98.1% and a specificity of 36.5%. A 60 mL distension had a significantly lower detection rate than 40 mL and 50 mL for active peristalsis and was unaffected by pressure (P < 0.05).

Conclusions: FLIP as an adjunct to HRIM is supported by strong metric correlation. FLIP was not correlated to pH monitoring findings, suggesting FLIP is not useful in reflux assessment. The agreement between FLIP and TBE was lower than in previous studies. Hiatal hernia impacted the normality between DI and IRP, not between FLIP and TBE. We suggest analyzing peristaltic patterns on panometry at all fill volumes to optimize detection.