A comparison of the usage of an open-source automated insulin delivery system and the MiniMed™ 780 G system in children and adolescents with type 1 diabetes in real-world settings: the AWeSoMe study group

Endocrine. 2024 Jan 16. doi: 10.1007/s12020-024-03683-w. Online ahead of print.

Abstract

Purpose: In recent years there has been a noticeable increase in the use of advanced hybrid closed-loop systems (AHCLs) for managing type 1 diabetes (T1D) among youth. However, there is a lack of comparison between the open-source automated insulin delivery (AID) system and the MiniMed™ 780 G system (780 G).

Methods: In this multi-center study, we retrospectively compared selected glycemic ranges of 26 individuals who used open-source AID and 20 individuals who used 780 G (age 11.3 years [IQR 9.3, 12.9] and 13.4 years [IQR 10.9, 16.5], respectively, p = 0.069) from system initiation to the most recent visit.

Results: At baseline, the median HbA1c was significantly lower and the time below range (TBR)<54mg/dL was significantly higher in the open-source AID group compared to the 780 G group (6.8% [IQR 6.4, 7.1] vs. 7.4% [IQR 6.9, 8.6], p = 0.006 and (1.0% [IQR 0.5, 2.8] vs. 0.0% [0.0, 1.0], p = 0.014), respectively; the median time in range (TIR70-180mg/dL) was similar (p = 0.068). After a median duration of 10.9 months on AHCLs the reduction of HbA1c was similar ( ~ 0.3%). The time spent in the hypoglycemic ranges was longer among users of the open-source AID compared to 780 G (TBR54-70mg/dL 4.2% [IQR 2.6, 7.3] vs. 2.0% [1.0, 4.0], p = 0.005) and TBR<54mg/dL 1.1% [IQR 0.4, 2.3] vs. 0.0 [0.0, 1.0], p = 0.001).

Conclusions: Both AHCLs similarly improved HbA1c and TIR70-180mg/dL. The open-source AID youth had better glycemic control but spent longer time in the hypoglycemic range. These findings must be considered when choosing the use of AHCL technologies.

Keywords: Advanced hybrid closed-loop systems (AHCL); Continuous glucose monitoring (CGM); Glycemic metrics; Type 1 diabetes.