A novel discrete linkage-type electrode for radiofrequency-induced intestinal anastomosis

Minim Invasive Ther Allied Technol. 2024 Apr;33(2):71-79. doi: 10.1080/13645706.2023.2291439. Epub 2024 Jan 14.

Abstract

Introduction: For decades, radiofrequency (RF)-induced tissue fusion has garnered great attention due to its potential to replace sutures and staples for anastomosis of tissue reconstruction. However, the complexities of achieving high bonding strength and reducing excessive thermal damage present substantial limitations of existing fusion devices.

Materials and methods: This study proposed a discrete linkage-type electrode to carry out ex vivo RF-induced intestinal anastomosis experiments. The anastomotic strength was examined by burst pressure and shear strength test. The degree of thermal damage was monitored through an infrared thermal imager. And the anastomotic stoma fused by the electrode was further investigated through histopathological and ultrastructural observation.

Results: The burst pressure and shear strength of anastomotic tissue can reach 62.2 ± 3.08 mmHg and 8.73 ± 1.11N, respectively, when the pressure, power and duration are 995 kPa, 160 W and 13 s, and the thermal damage can be controlled within limits. Histopathological and ultrastructural observation indicate that an intact and fully fused stomas with collagenic crosslink can be formed.

Conclusion: The discrete linkage-type electrode presents favorable efficiency and security in RF-induced tissue fusion, and these results are informative to the design of electrosurgical medical devices with controllable pressure and energy delivery.

Keywords: Radiofrequency energy; anastomotic strength; electrode design; thermal damage; tissue fusion.

MeSH terms

  • Anastomosis, Surgical / methods
  • Collagen
  • Digestive System Surgical Procedures*
  • Electrodes

Substances

  • Collagen