Decreasing implant load indicates spinal fusion when measured continuously

J Biomech. 2024 Jan:163:111929. doi: 10.1016/j.jbiomech.2024.111929. Epub 2024 Jan 10.

Abstract

Reliable and timely assessment of bone union between vertebrae is considered a key challenge after spinal fusion surgery. Recently, a novel sensor concept demonstrated the ability to objectively assess posterolateral fusion based on continuous implant load monitoring. The aim of this study was to investigate systematically the concept in a mono-segmental fusion model using an updated sensor setup. Three sheep underwent bilateral facetectomy at level L2-L3 and L4-L5. The segments were stabilized using two unconnected pedicle-screw-rod constructs per level. Sensing devices were attached to the rods between each pedicle screw pair and the loads were continuously monitored over 16 weeks. After euthanasia, the spines were biomechanically tested for their range of motion and high-resolution CT scans were performed to confirm the fusion success. After an initial increase in implant load until reaching a maximum (100 %) at approximately week 4, eleven out of twelve sensors measured a constant decrease in implant load to 52 ± 9 % at euthanasia. One sensor measurement was compromised by newly forming bone growing against the sensor clamp. Bridging bone at each facet and minor remnant segmental motion (<0.7°) confirmed the fusion of all motion segments. Data obtained by continuous measurement of implant loading of spinal screw-rod constructs enables objective monitoring of spinal fusion progression. The sensor concept provides valuable real-time information, offering quantifiable data as an alternative to traditional imaging techniques. However, the design of the current sensor concept needs to be matured, tailored to, and validated for the human spine.

Keywords: Continuous measurement; Fusion assessment; Implant load; In vivo measurements; Spinal fusion.

MeSH terms

  • Animals
  • Biomechanical Phenomena
  • Humans
  • Lumbar Vertebrae / surgery
  • Pedicle Screws*
  • Range of Motion, Articular
  • Sheep
  • Spinal Fusion*