Introduction of a novel neonatal warming device in Malawi: an implementation science study

Int Health. 2024 Jan 12:ihad114. doi: 10.1093/inthealth/ihad114. Online ahead of print.

Abstract

Background: Neonatal hypothermia significantly contributes to infant morbidity and mortality in low-resource settings like Malawi. Kangaroo mother care (KMC) is essential but faces challenges in providing continuous thermal support. The Dream Warmer is a neonatal warming device that was developed to complement KMC. We studied its implementation outside a research environment.

Methods: Using an implementation science approach, we conducted a prospective interventional cohort study in two hospitals and four health centres in Malawi. Through audits and surveys, we assessed the effect of the Dream Warmer on neonatal hypothermia as well as healthcare provider (HCP) and parent attitudes regarding thermoregulation and related issues.

Results: The Dream Warmer raised no safety concerns and effectively treated hypothermia in 90% of uses. It was positively received by HCPs and parents, who reported it had a favourable effect on the care of small and sick newborns. Challenges identified included a scarcity of water and electricity, lack of availability of the device and HCPs forgetting to prepare it in advance of need or to use it when indicated. Feedback for future training was obtained. The Dream Warmer's strong safety and effectiveness performance is consistent with results from strict research studies. Training materials can be adapted to optimize integration into daily practice and provide educational content for parents.

Conclusions: The Dream Warmer is a safe and effective device to treat neonatal hypothermia, particularly when KMC is insufficient. We gained an understanding of how to optimize implementation through robust HCP and family education to help combat hypothermia.

Keywords: KMC; LMIC; Malawi; implementation science; neonatal hypothermia; newborn.