A comparative trial of blood pressure monitoring in a self-care kiosk, in office, and with ambulatory blood pressure monitoring

BMC Cardiovasc Disord. 2024 Jan 3;24(1):27. doi: 10.1186/s12872-023-03701-1.

Abstract

Background: Automated measurement of blood pressure (BP) in designated BP kiosks have in recent years been introduced in primary care. If kiosk blood pressure (BP) monitoring provides results equivalent to in-office BP or daytime ambulatory BP monitoring (ABPM), follow-up of adult patients could be managed primarily by self-checks. We therefore designed a comparative trial and evaluated the diagnostic performance of kiosk- and office-based BP (nurse- versus physician-measured) compared with daytime ABPM.

Methods: A trial of automated BP monitoring in three settings: a designated BP kiosk, by nurses and physicians in clinic, and by ABPM. The primary outcome was systolic and diastolic BP, with respective diagnostic thresholds of ≥135 mmHg and/or ≥ 85 mmHg for daytime ABPM and kiosk BP and ≥ 140 mmHg and/or ≥ 90 mmHg for office BP (nurse- and physician-measured).

Results: Compared with daytime ABPM, mean systolic kiosk BP was higher by 6.2 mmHg (95% confidence interval [CI] 3.8-8.6) and diastolic by 7.9 mmHg (95% CI 6.2-9.6; p < 0.001). Mean systolic BP taken by nurses was similar to daytime ABPM values (+ 2.0 mmHg; 95% CI - 0.2-4.2; p = 0.071), but nurse-measured diastolic values were higher, by 7.2 mmHg (95% CI 5.9-9.6; p < 0.001). Mean systolic and diastolic physician-measured BPs were higher compared with daytime ABPM (systolic, by 7.6 mmHg [95% CI 4.5-10.2] and diastolic by 5.8 mmHg [95% CI 4.1-7.6]; p < 0.001). Receiver operating characteristic curves of BP monitoring across pairs of systolic/diastolic cut-off levels among the three settings, with daytime ABPM as reference, demonstrated overall similar diagnostic performance between kiosk and nurse-measured values and over the curve performance for physician-measured BP. Accuracy with nurse-measured BP was 69.2% (95% CI 60.0-77.4%), compared with 65.8% (95% CI 56.5-74.3%) for kiosk BP.

Conclusions: In this study kiosk BP monitoring was not comparable to daytime ABPM but could be an alternative to in-office BP monitoring by trained nurses. The diagnostic performance of kiosk and nurse-measured BP monitoring was similar and better than that of physician-measured BP.

Trial registration: ClinicalTrials.gov (NCT04488289) 27/07/2020.

Keywords: Ambulatory blood pressure; Clinical trial; Self-blood pressure monitoring.

Publication types

  • Clinical Trial
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Blood Pressure
  • Blood Pressure Determination / methods
  • Blood Pressure Monitoring, Ambulatory*
  • Humans
  • Hypertension* / diagnosis
  • Self Care

Associated data

  • ClinicalTrials.gov/NCT04488289