Background: The aim of this study was to compare shoulder movement measurements between a Kinect-based markerless ROM assessment device (POM-Checker) and a 3D motion capture analysis system (BTS SMART DX-400).
Methods: This was a single-visit clinical trial designed to evaluate the validity and reliability of the POM-Checker. The primary outcome was to assess the equivalence between two measurement devices within the same set of participants, aiming to evaluate the validity of the POM-Checker compared to the gold standard device (3D Motion Analysis System). As this was a pilot study, six participants were included.
Results: The intraclass correlation coefficient (ICC) and the corresponding 95% confidence intervals (CIs) were used to assess the reproducibility of the measurements. Among the 18 movements analyzed, 16 exhibited ICC values of >0.75, indicating excellent reproducibility.
Conclusion: The results showed that the POM-checker is reliable and validated to measure the range of motion of the shoulder joint.
Keywords: comparative study; pilot study; range of motion; shoulder joint.