The Dose Optimization and Evaluation of Image Quality in the Adult Brain Protocols of Multi-Slice Computed Tomography: A Phantom Study

J Imaging. 2023 Nov 28;9(12):264. doi: 10.3390/jimaging9120264.

Abstract

Computed tomography examinations have caused high radiation doses for patients, especially for CT scans of the brain. This study aimed to optimize the radiation dose and image quality in adult brain CT protocols. Images were acquired using a Catphan 700 phantom. Radiation doses were recorded as CTDIvol and dose length product (DLP). CT brain protocols were optimized by varying parameters such as kVp, mAs, signal-to-noise ratio (SNR) level, and Clearview iterative reconstruction (IR). The image quality was also evaluated using AutoQA Plus v.1.8.7.0 software. CT number accuracy and linearity had a robust positive correlation with the linear attenuation coefficient (µ) and showed more inaccurate CT numbers when using 80 kVp. The modulation transfer function (MTF) showed a higher value in 100 and 120 kVp protocols (p < 0.001), while high-contrast spatial resolution showed a higher value in 80 and 100 kVp protocols (p < 0.001). Low-contrast detectability and the contrast-to-noise ratio (CNR) tended to increase when using high mAs, SNR, and the Clearview IR protocol. Noise decreased when using a high radiation dose and a high percentage of Clearview IR. CTDIvol and DLP were increased with increasing kVp, mAs, and SNR levels, while the increasing percentage of Clearview did not affect the radiation dose. Optimized protocols, including radiation dose and image quality, should be evaluated to preserve diagnostic capability. The recommended parameter settings include kVp set between 100 and 120 kVp, mAs ranging from 200 to 300 mAs, SNR level within the range of 0.7-1.0, and an iterative reconstruction value of 30% Clearview to 60% or higher.

Keywords: dose optimization; image quality; iterative reconstruction; radiation dose.

Grants and funding

This research received no external funding.