Development of Biopsy Tract Sealants Based on Shape Memory Polymer Foams

Biomed Mater Devices. 2023 Sep;1(2):853-860. doi: 10.1007/s44174-023-00064-1. Epub 2023 Feb 8.

Abstract

Lung biopsies are often used to aid in the diagnosis of cancers. However, the procedure carries the dual risk of air (pneumothorax) or blood (hemothorax) filling the pleural cavity, increasing the risk of a collapsed lung and chest intubation. This work demonstrates the effectiveness of a polyurethane-based shape memory polymer foam as a biopsy tract sealant. The impact of diameter, length, pore size, and shape memory effect was evaluated to determine the ideal device design for tract sealing. Characterization in an in vitro benchtop lung model identified that diameter had the largest influence on sealing efficacy, while the length of the device had little to no impact. Finally, evaluation of deployment force demonstrated that devices fabricated from the shape memory polymer foams were easier to deploy than elastic foams. Following characterization, down-selected device designs were combined with radiopaque markers for use in image-guided based procedures. Furthermore, the introduction of the markers or sterilization did not impact the ability of the devices to seal the biopsy tract and led to a decrease in the deployment force. Overall, these results demonstrate the potential for polyurethane-based shape memory foam devices to serve as biopsy tract sealant devices that aim to reduce complications, such as pneumothorax, from occurring.

Keywords: Lung biopsy; pneumothorax; sealant; shape memory polymer foam.