Comparing the diagnostic accuracy, reading time, and inter-rater agreement of breast MRI abbreviated and full protocols: a multi-reader study

Acta Radiol. 2024 Feb;65(2):195-201. doi: 10.1177/02841851231216552. Epub 2023 Dec 19.

Abstract

Background: Earlier studies have shown that abbreviated protocol magnetic resonance imaging (AB-MRI) has similar diagnostic accuracy as the full protocol (Full MRI).

Purpose: To compare the diagnostic accuracy, reading time, and inter-rater agreement of AB-MRI to Full MRI among women without known increased familial risk of breast cancer or prior biopsy.

Material and methods: In total, 395 MRI examinations were included in this study. Three readers were blinded to all patient information. The AB-MRI and Full MRI were read separately and in a different random order for each of the readers. Scores 1-2 were considered test negative while scores 3-5 were test positive. A positive reference test was the diagnosis of malignancy; a negative reference test was the absence of a diagnosis of breast cancer within a two-year follow-up. We used a generalized estimating equations approach to compare sensitivity and specificity between the two protocols. We used t-tests to compare the average reading time and Krippendorff's alpha to compare inter-rater agreement.

Results: MRI examinations of 395 women (median age=56 years) were evaluated. For AB-MRI and Full MRI, respectively, the sensitivity was 93.0% (95% CI=90.6-95.0) vs. 92.0% (95% CI=89.4-94.1), the specificity was 91.7% (95% CI=90.3-92.9) vs. 94.3% (95% CI=93.2-95.3), average reading time was 67 vs. 126 s, and the inter-rater agreement 0.79 vs. 0.83. The difference in sensitivity was not statistically significant (P=0.840), but the difference in specificity was significant (P=0.003).

Conclusion: AB-MRI has similar sensitivity, but somewhat lower specificity. The average reading time for the abbreviated protocol is lower, as is inter-rater agreement.

Keywords: Breast; magnetic resonance imaging; neoplasm.

MeSH terms

  • Breast Neoplasms* / diagnostic imaging
  • Breast Neoplasms* / pathology
  • Breast* / diagnostic imaging
  • Breast* / pathology
  • Female
  • Humans
  • Magnetic Resonance Imaging / methods
  • Middle Aged
  • Radiography
  • Retrospective Studies
  • Sensitivity and Specificity