Simultaneous determination of 4-ethylphenol and 4-ethylguaicol on C60 modified dual screen-printed electrochemical sensors

Talanta. 2024 Apr 1:270:125543. doi: 10.1016/j.talanta.2023.125543. Epub 2023 Dec 13.

Abstract

4-ethylphenol and 4-ethylguaicol levels in wine are associated to organoleptic defects that cause consumer rejection accompanied by significant economic losses for producers. Thus, electrochemical sensors based on screen-printed carbon electrodes (SPCEs) modified with activated fullerene C60 (AC60) have been developed for the analysis of both phenols by direct headspace amperometric measurements. Upon optimization of the experimental variables affecting the sensors performance, the AC60/SPCE sensors presented linearity ranges from 9.9 to 65.4 μg/L and from 19.6 to 107.1 μg/L for 4-ethylphenol and 4-ethylguaicol, respectively. The achieved detection capacities were 10.3 μg/L (4-ethylphenol) and 19.6 μg/L (4-ethylguaicol), with a reproducibility of 6.3 % and 9.1 % (n = 3), respectively. In addition, dual-working AC60/SPCE devices were developed for the simultaneous analysis of both phenols using different working potentials for each electrode. The dual systems were successfully applied in the analysis of different spiked wine samples, obtaining good recoveries ranging from 94 to 108 %.

Keywords: 4-Ethylguaicol; 4-Ethylphenol; Dual screen-printed electrodes; Fullerene C(60); Wine.