Abstract

Remote camera-based estimation of blood oxygen saturation (SpO2) using visible lights has been studied recently, typically for red (660 nm) and green (550 nm) wavelengths. This paper investigates the impact of different skin penetration depths of red and green wavelengths on the SpO2 estimation based on mathematical modeling and experiments, where the SpO2-calibritability between two illumination setups, narrow-band red/green and narrow-band red/infrared, are statistically compared using the "ratio-of-ratios" method. The results show that the performance of the setup using red/green is less consistent among 17 volunteers than the setup using red/infrared, and larger SpO2 disparity between different skin regions (by SpO2 imaging) have been found for individuals in the red/green wavelengths setup. The use of visible light (red and green) may impose a risk of SpO2 calibration due to the different skin penetration depths of these two wavelengths.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Humans
  • Light
  • Oximetry / methods
  • Oxygen Saturation*
  • Photoplethysmography* / methods
  • Pulmonary Gas Exchange