MRI compatibility testing of commercial high intensity focused ultrasound transducers

Phys Med. 2024 Jan:117:103194. doi: 10.1016/j.ejmp.2023.103194. Epub 2023 Dec 3.

Abstract

Purpose: The study aimed to compare the performance of eight commercially available single-element High Intensity Focused Ultrasound (HIFU) transducers in terms of Magnetic Resonance Imaging (MRI) compatibility.

Methods: Imaging of an agar-based MRI phantom was performed in a 3 T MRI scanner utilizing T2-Weighted Fast Spin Echo (FSE) and Fast low angle shot (FLASH) sequences, which are typically employed for high resolution anatomical imaging and thermometry, respectively. Reference magnitude and phase images of the phantom were compared with images acquired in the presence of each transducer in terms of the signal to noise ratio (SNR), introduced artifacts, and overall image quality.

Results: The degree of observed artifacts highly differed among the various transducers. The transducer whose backing material included magnetic impurities showed poor performance in the MRI, introducing significant susceptibility artifacts such as geometric distortions and signal void bands. Additionally, it caused the most significant SNR drop. Other transducers were shown to exhibit high level of MRI compatibility as the resulting images closely resembled the reference images with minimal to no apparent artifacts and comparable SNR values.

Conclusions: The study findings may facilitate researchers to select the most suitable transducer for their research, simultaneously avoiding unnecessary testing. The study further provides useful design considerations for MRI compatible transducers.

Keywords: Artifacts; Commercial; Compatibility; HIFU; MRI; Transducer.

MeSH terms

  • Artifacts
  • Magnetic Resonance Imaging* / methods
  • Phantoms, Imaging
  • Signal-To-Noise Ratio
  • Transducers*