Analysis of Return-to-Zero Error after the First Load of Load Cell

Sensors (Basel). 2023 Oct 25;23(21):8712. doi: 10.3390/s23218712.

Abstract

The return-to-zero error of the resistance strain load cell is most obvious in the first zero-return process during loading and unloading. To improve the accuracy of the load cell, it is necessary to figure out the cause of the error. The influence of the temperature, material, and weld cup were analyzed in this paper. It was concluded that the hysteresis is the main factor affecting the return-to-zero error after the first load. The relationship between hysteresis and zero-return error after first load was obtained by a data fitting algorithm. A method to improve the return-to-zero error after the first load was proposed.

Keywords: Pearson correlation coefficient; data fitting algorithm; hysteresis; zero-point error.

Grants and funding

This research received no external funding.