A practical method to simulate realistic reduced-exposure CT images by the addition of computationally generated noise

Radiol Phys Technol. 2024 Mar;17(1):112-123. doi: 10.1007/s12194-023-00755-w. Epub 2023 Nov 13.

Abstract

Computed tomography (CT) scanning protocols should be optimized to minimize the radiation dose necessary for imaging. The addition of computationally generated noise to the CT images facilitates dose reduction. The objective of this study was to develop a noise addition method that reproduces the complexity of the noise texture present in clinical images with directionality that varies over images according to the underlying anatomy, requiring only Digital Imaging and Communications in Medicine (DICOM) images as input data and commonly available phantoms for calibration. The developed method is based on the estimation of projection data by forward projection from images, the addition of Poisson noise, and the reconstruction of new images. The method was validated by applying it to images acquired from cylindrical and thoracic phantoms using source images with exposures up to 49 mAs and target images between 39 and 5 mAs. 2D noise spectra were derived for regions of interest in the generated low-dose images and compared with those from the scanner-acquired low-dose images. The root mean square difference between the standard deviations of noise was 4%, except for very low exposures in peripheral regions of the cylindrical phantom. The noise spectra from the corresponding regions of interest exhibited remarkable agreement, indicating that the complex nature of the noise was reproduced. A practical method for adding noise to CT images was presented, and the magnitudes of noise and spectral content were validated. This method may be used to optimize CT imaging.

Keywords: Low-dose simulation; Noise modelling; Radiation dose reduction; Two-dimensional noise spectra; X-ray computed tomography.

MeSH terms

  • Algorithms*
  • Calibration
  • Phantoms, Imaging
  • Radiation Dosage
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed* / methods