Continuous glucose monitoring (CGM) is proposed as an alternative for glycemic assessment in peritoneal dialysis, but volume overload and anemia may affect sensor accuracy. This is an exploratory analysis of a study of Guardian Connect™ with Guardian Sensor™ 3 in 30 participants with diabetes on continuous ambulatory peritoneal dialysis (CAPD) (age [mean ± standard deviation] 64.7 ± 5.6 years, 23 men, body mass index [BMI] 25.4 ± 3.9 kg/m2, blood hemoglobin [Hb] 10.7 ± 1.3 g/dL). The mean absolute relative difference (MARD) was calculated between paired sensor and YSI 2300 STAT venous glucose readings (n = 941) during an 8-h in-clinic session with glucose challenge. Body composition was evaluated using bioimpedance. The overall MARD was 10.4% (95% confidence interval 9.6-11.7). There were no correlations between BMI, extracellular water, relative hydration index, and lean or fat mass with MARD. No correlations were observed between MARD and Hb (r = 0.016, P > 0.05). In summary, this real-time CGM demonstrated good accuracy in CAPD with minimal influence from body composition and anemia.
Keywords: Body composition; Continuous glucose monitoring; Diabetic kidney disease; Peritoneal dialysis.