Intraindividual Comparison of an Enhanced Monofocal and an Aspheric Monofocal Intraocular Lens of the Same Platform

Am J Ophthalmol. 2024 May:261:95-102. doi: 10.1016/j.ajo.2023.11.006. Epub 2023 Nov 8.

Abstract

Purpose: To compare intraindividual differences in visual performance of a monofocal and enhanced monofocal intraocular lens (IOL) of the same platform.

Design: Prospective, interventional, fellow-eye comparison clinical study.

Methods: In total, 55 patients (110 eyes) with bilateral age-related cataract were enrolled. All patients received a monofocal ZCB00 IOL in the dominant and an enhanced monofocal Eyhance ICB00 IOL in the nondominant eye. After 2 to 4 months, monocular best-corrected distant visual acuity (BCDVA), distance-corrected intermediate visual acuity (DCIVA), distance-corrected near visual acuity (DCNVA), contrast visual acuity, monocular defocus curves, internal higher-order aberrations (HOAs) and spherical aberrations (SA), decentration, and tilt were compared.

Results: The monocular mean BCDVA, DCIVA at 80 cm and 66 cm, and DCNVA were -0.03 ± 0.07, 0.24 ± 0.12, 0.32 ± 0.13, and 0.50 ± 0.13 logarithm of the minimum angle of resolution for the enhanced ICB00 and -0.06 ± 0.06 (P = .014), 0.30 ± 0.11 (P = .005), 0.38 ± 0.12 (P = .004), and 0.55 ± 0.14 (P = .034) logarithm of the minimum angle of resolution for the ZCB00, respectively. Internal HOAs (P = .001) and negative SA (P < .001) were increased with the ICB00 at 3 mm and comparable at 5 mm (P > .05). Contrast acuity, tilt, and decentration were similar (P > .05).

Conclusions: Significantly increased monocular DCIVA at 80 cm and 66 cm and DCNVA at 40 cm were observed with the enhanced ICB00 IOL, and the ZCB00 IOL demonstrated better BCDVA. This would result in a mean gain of 2 to 3 Early Treatment of Diabetic Retinopathy Study letters at near and intermediate distance. Monocular defocus curves displayed highest differences of 5 Early Treatment of Diabetic Retinopathy Study letters at -1.25 diopters (D) and -1.50 D levels of defocus and a depth of focus of 1.23 D for the ICB00 IOL and 0.94 D for the ZCB00 IOL. Decentration, tilt, and HOAs were generally low.