Simulation of double resonant excitation of ions in an asymmetric linear ion trap mass analyzer

Rapid Commun Mass Spectrom. 2023 Dec 15;37(23):e9625. doi: 10.1002/rcm.9625.

Abstract

Rationale: Improving the analytical performance of linear ion traps (LITs) is crucial for the advancement of high-performance LIT mass spectrometers. In this study, a double resonant excitation method was employed in an asymmetric LIT to achieve high ion unidirectional ejection efficiency and enhanced mass resolution.

Methods: The asymmetric trapping field was generated by stretching one x electrode with a distance α. The double resonant excitation was achieved by applying an alternating voltage out of phase and a supplementary alternating voltage in phase to the x and y electrode pairs of the LIT, respectively. Numerical simulations of ion trajectories were performed to validate the effectiveness of this method.

Results: The mass resolution of the asymmetric LIT with double resonant excitation could be improved to ~3800, which was over two times compared to that with only dipolar resonant excitation, while both reached ~90% in ion unidirectional ejection efficiency.

Conclusions: By employing the double resonant excitation method, the mass resolution could be improved significantly in the asymmetric LIT, while maintaining a considerably high ion unidirectional ejection efficiency. This method might provide a general solution for enhancing ion detection efficiency and mass resolution of LIT mass spectrometers.